AnalytikJena's Model TL-2000 UV Translinker includes the shortwave UV CL-1000 Crosslinker and an M-20 type transilluminator built onto the crosslinker cabinet. It supplies uniform intensity overhead UV exposure of materials inside the chamber and uses midrange 302nm UV for transillumination. The shortwave ultraviolet energy bonds DNA to a medium. Ultraviolet Crosslinking for attaching nucleic acids to a membrane takes seconds as compared to oven baking.

AnalytikJena Crosslinkers offer researchers an instrument to quickly, safely and efficiently expose samples to a controlled amount of ultraviolet radiation. A unique UV sensor continually measures the UV energy and automatically adjusts to variations in UV intensity that occurs as the UV tubes age. This same UV sensor feedback measurement system allows you to set UV sample exposure, which automatically deactivates the UV sources when the set UV energy dose has been achieved.

Applications:

Exposure of samples to UV provides for the following processes:

  • Crosslinking of DNA or RNA to nitrocellulose, nylon or reinforced nitrocellulose,
  • PCR sample contamination control,
  • Photoniking of DNA,
  • Testing RNA function,
  • Rapid site mapping,
  • UV sterilization and sanitization,
  • UV curing.

Features:

  • Preset and manual controls for ultraviolet or time exposures:
    • Factory presettings deliver 120,000 microjoules or 5 minutes of exposure,
    • Maximum UV energy setting of 999,900 microjoules per cm2.
  • Integrated UV sensor for measurement and exposure control is calibrated at the UVP factory to ensure accuracy and reliability,
  • Microprocessor output control optimizes the use of energy while ensuring consistent, maximum energy efficiency and output,
  • Large LED display with touch panel displays time or energy settings,
  • On/Off controls for the transilluminator are located on the crosslinker panel,
  • Window on the door allows viewing of the process yet blocks the UV radiation,  
  • Small footprint for benchtop use,
  • Five 8-watt ultraviolet tubes provide uniform overhead ultraviolet illumination,
  • Transilluminator has a 20 x 20cm active area with midrange 302nm UVB illumination,
  • An ultraviolet blocking cover shields the user from UV radiation.

Specifications:

  • Crosslinker:
    • Display: LED
    • Controls: Touch pad
    • Watts: 8 watt
    • UV exposure output: 120,000 microjoules or 5 minutes (adjustable); maximum UV energy setting of 999,900 microjoules per cm2
    • UV Source: 254nm
    • No. of tubes: 5 bi-pin type tubes
  • Transilluminator:
    • UV Source: 302nm
    • Watts: 8 watt
    • Filter size: 20 x 20cm
    • UV blocking cover included
  • TL-2000 Dimensions:
    • External: 13.7D x 19W x 12.75H in. (348 x 483 x 323mm)
    • Internal: 12D x 10W x 5H in. (305 x 254 x 127mm)
    • Weight: 36 lb (16.4 kg)
  • Select input Voltage:  

Part Numbers

Volts

95-0300-01 115V
95-0300-02 230V

 

Optional Accessories:

---

Warranty Information

AnalytikJena 's products are guaranteed to be free of defects in materials, workmanship and manufacture for one (1) year from date of purchase, except as follows:
- Consumable and disposable parts are guaranteed to be free from defects in manufacture and materials for ninety (90) days from date of purchase.
- UV tubes and filters are guaranteed to be free from defects in manufacture and materials for ninety (90) days from date of purchase.
- AnalytikJena Transilluminators are guaranteed to be free from defects in manufacture and materials for two (2) years from date of purchase.

If equipment failure or malfunction occurs during the warranty period, AnalytikJena shall examine the inoperative equipment and have the option of repairing or replacing any part(s) which, in the judgment of AnalytikJena , were originally defective or became so under conditions of normal usage and service.

No warranty shall apply to this instrument, or part thereof, that has been subject to accident, negligence, alteration, abuse or misuse by the end-user. Moreover, AnalytikJena makes no warranties whatsoever with respect to parts not supplied by AnalytikJena or that have been installed, used and/or serviced other than in strict compliance with instructions appearing in this manual.

In no event shall AnalytikJena be responsible to the end-user for any incidental or consequential damages, whether foreseeable or not, including but not limited to property damage, inability to use equipment, lost business, lost profits, or inconvenience arising out of or connected with the use of instruments produced by AnalytikJena . Nor is AnalytikJena liable or responsible for any personal injuries occurring as a result of the use, installation and/or servicing of equipment.

This warranty does not supersede any statutory rights that may be available in certain countries.

Find Similar Products by Category

Write your own product review

Product Reviews

This product hasn't received any reviews yet. Be the first to review this product!